【Maintenance Case】Failure to Connect Flat Panel Detector of Newtom Series CBCT Weiqiu Professional Maintenance: X-ray tube repair, complete machine warranty & technical support
2026-06-150 ReadDR / DSA / C-arm / CT / Gastrointestinal X-ray machine

Device Brand: Newtom, Italy
Device Model: Newtom VGI CBCT
Fault Phenomenon: Failed to connect the detector
Device Fault Photos:

1. Troubleshooting: The flat panel detector fails to connect. Inspection shows the network cable for the detector is faulty and needs replacement.


2. Maintenance process: a. Replace the dedicated flexible network cable


2. Maintenance process: b. The on-board computer was heavily soiled. Clean the computer and equipment cabinet.


3. Maintenance completed: The device starts up, moves and exposes normally. Image acquisition and data transmission work properly. The equipment is fully operational now!